How the XF Extracellular Flux Analyzer Works
From the beginning, Seahorse Bioscience has worked closely with senior scientists, directors and distinguished research fellows at some of the world's largest pharmaceutical companies including Abbott Laboratories, Amgen, AstraZeneca, Bristol Myers Squibb, GlaxoSmithKline, Harvard, Johnson & Johnson, and Merck & Company.
The vision was to develop a series of non-invasive instruments that enabled scientists to measure cellular bioenergetics in a microplate format - quickly, easily and in real-time.
A testament to the fruition of that vision and utility of the XF Series is the number of different therapeutic groups within each organization that are leveraging this novel technology today.
How Extracellular Flux is Measured
The XF Analyzer measures oxygen consumption rate (OCR) and extracellular acidification rate (ECAR) at intervals of approximately 2-5 minutes. OCR is an indicator of mitochondrial respiration, and ECAR is predominately the result of glycolysis.
Real time measurements of OCR and ECAR are made by isolating an extremely small volume (less than 7ul) of medium above a monolayer of cells within a microplate. Cellular oxygen consumption (respiration) and proton excretion (glycolysis) causes rapid, easily measurable changes to the concentrations of dissolved oxygen and free protons in this "transient microenvironment" which are measured every few seconds by solid state sensor probes residing 200 microns above the cell monolayer.
The instrument continues to measure the concentrations until the rate of change is linear -- usually 10% for O2, and 0.2 mpH units for H+ -- and then calculates the slope to determine OCR and ECAR, respectively.
Once a measurement is completed, the probes lift which allows the larger medium above to mix with the medium in the transient microchamber, restoring cell values to baseline. Measurements may be repeated indefinitely since no dyes are added.
An integrated drug delivery system allows sequential addition of up to four drugs per well at user-defined intervals.
How an XF Assay is Performed
Prior to the start of the XF assay, cells are seeded into the wells of an XF cell culture microplate. Seahorse supplied, bicarbonate free medium is used during the assay. Up to two or four drugs may be added to the assay cartridge for use during the assay.
The assay cartridge is first placed into the XF Analyzer to allow automatic calibration of optical sensors. Next, the cell culture plate is inserted into the instrument.
Typically, baseline (basal) OCR and ECAR rates are measured twice before drug addition. OCR is reported in units of nmoles/minute and ECAR in mpH/minute. Drugs that have been preloaded into the drug delivery chambers of the assay cartridge are then pneumatically injected into the media in each well. After mixing, post-exposure OCR and ECAR measurements are made multiple times.
Because XF measurements are non-destructive, the metabolic rate of the same cell population can be measured repeatedly over time while up to two or four different drugs can be injected sequentially into each well. Upon completion of an XF assay, other types of biological assays such as cell viability can be performed on the same plate.
Total assay time is typically 35 to 90 minutes. To maintain normal cell physiology, a temperature control system maintains the XF Analyzer's internal environment at 37 degrees Celsius.
Instrument Sensitivity: The Results Speak for Themselves.
As a result of a series of novel inventions, we believe that our technology has exceeded the performance of any previous non-invasive metabolic readout. Measurements of cellular metabolism, from one well to the next, one plate to the next, or one day to the next, typically produce c.v.'s of less than 20%. Measurements of the change in cellular behavior as a result of drug addition typically produce c.v.'s of less than 10%. Changes of as little as 5% can often be seen with replicate wells.